Performance of a low gain avalanche detector in a medical linac and characterisation of the beam profile.
Saved in:
| Title: | Performance of a low gain avalanche detector in a medical linac and characterisation of the beam profile. |
|---|---|
| Authors: | Isidori, T (AUTHOR) ronan.mcnulty@ucd.ie, McCavana, P (AUTHOR), McClean, B (AUTHOR), McNulty, R (AUTHOR), Minafra, N (AUTHOR), Raab, N (AUTHOR), Rock, L (AUTHOR), Royon, C (AUTHOR) |
| Source: | Physics in Medicine & Biology. 7/7/2021, Vol. 66 Issue 13, p1-10. 10p. |
| Subjects: | Radiation dosimetry, Detectors, Radiation measurements, Linear accelerators, Picosecond pulses |
| Abstract: | Low gain avalanche detectors can measure charged particle fluences with high speed and spatial precision, and are a promising technology for radiation monitoring and dosimetry. A detector has been tested in a medical linac where single particles were observed with a time resolution of 50 ps. The integrated response is similar to a standard ionising chamber but with a spatial precision twenty times finer, and a temporal precision over 100 million times better, with the capability to measure the charge deposited by a single linac pulse. The unprecedented resolving power allows the structure of the ∼3 μs linac pulses to be viewed and the 350 ps sub-pulses in the train to be observed. [ABSTRACT FROM AUTHOR] |
| © 2025 IOP Publishing Ltd. All rights, including for text and data mining, AI training, and similar technologies, are reserved (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |