A New Rate-Modulated Pacemaker System Optimized by Combination of Two Sensors.

Saved in:
Bibliographic Details
Title: A New Rate-Modulated Pacemaker System Optimized by Combination of Two Sensors.
Authors: Alt, Eckhard1, Theres, Heinz1, Heinz, Michael1, Matula, Markus1, Thilo, Rudolf1, Blömer, Hans1
Source: Pacing & Clinical Electrophysiology. Aug1988, Vol. 11 Issue 8, p1119-1129. 11p.
Subjects: Cardiac pacemakers, Central venous pressure, Blood pressure, Exercise, Fever, Venous pressure
Abstract: A new rate-modulated pacemaker system optimized by combination of two sensors is described. The parameter body activity and central venous blood temperature control the pacemaker rate. The specific characteristic of each parameter determines its role within the algorithm. While the motion sensor yields a fast reaction following the onset OT a change of stress intensity, central venous blood temperature corresponds better to body metabolism. An indication of increased exercise from the motion sensor results in an accordingly rapid increase in the pacing rate. Unless this increased exercise is confirmed by an increase in central venous blood temperature within 2 or 3 minutes, the new motion level will be assumed to he the new baseline motion value and the pace rate will return to a basic pacing rate. Prolonged inappropriate responses are therefore avoided. Longer lasting exercise, fever and nonphysiological signals are recognized and handled safely. Exercise tests with five volunteers under various conditions showed pacing rate behavior that was close to normal. [ABSTRACT FROM AUTHOR]
Copyright of Pacing & Clinical Electrophysiology is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Database: Engineering Source
FullText Links:
  – Type: pdflink
Text:
  Availability: 0
Header DbId: egs
DbLabel: Engineering Source
An: 17484561
AccessLevel: 6
PubType: Academic Journal
PubTypeId: academicJournal
PreciseRelevancyScore: 0
IllustrationInfo
Items – Name: Title
  Label: Title
  Group: Ti
  Data: A New Rate-Modulated Pacemaker System Optimized by Combination of Two Sensors.
– Name: Author
  Label: Authors
  Group: Au
  Data: <searchLink fieldCode="AR" term="%22Alt%2C+Eckhard%22">Alt, Eckhard</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Theres%2C+Heinz%22">Theres, Heinz</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Heinz%2C+Michael%22">Heinz, Michael</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Matula%2C+Markus%22">Matula, Markus</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Thilo%2C+Rudolf%22">Thilo, Rudolf</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Blömer%2C+Hans%22">Blömer, Hans</searchLink><relatesTo>1</relatesTo>
– Name: TitleSource
  Label: Source
  Group: Src
  Data: <searchLink fieldCode="JN" term="%22Pacing+%26+Clinical+Electrophysiology%22">Pacing & Clinical Electrophysiology</searchLink>. Aug1988, Vol. 11 Issue 8, p1119-1129. 11p.
– Name: Subject
  Label: Subjects
  Group: Su
  Data: <searchLink fieldCode="DE" term="%22Cardiac+pacemakers%22">Cardiac pacemakers</searchLink><br /><searchLink fieldCode="DE" term="%22Central+venous+pressure%22">Central venous pressure</searchLink><br /><searchLink fieldCode="DE" term="%22Blood+pressure%22">Blood pressure</searchLink><br /><searchLink fieldCode="DE" term="%22Exercise%22">Exercise</searchLink><br /><searchLink fieldCode="DE" term="%22Fever%22">Fever</searchLink><br /><searchLink fieldCode="DE" term="%22Venous+pressure%22">Venous pressure</searchLink>
– Name: Abstract
  Label: Abstract
  Group: Ab
  Data: A new rate-modulated pacemaker system optimized by combination of two sensors is described. The parameter body activity and central venous blood temperature control the pacemaker rate. The specific characteristic of each parameter determines its role within the algorithm. While the motion sensor yields a fast reaction following the onset OT a change of stress intensity, central venous blood temperature corresponds better to body metabolism. An indication of increased exercise from the motion sensor results in an accordingly rapid increase in the pacing rate. Unless this increased exercise is confirmed by an increase in central venous blood temperature within 2 or 3 minutes, the new motion level will be assumed to he the new baseline motion value and the pace rate will return to a basic pacing rate. Prolonged inappropriate responses are therefore avoided. Longer lasting exercise, fever and nonphysiological signals are recognized and handled safely. Exercise tests with five volunteers under various conditions showed pacing rate behavior that was close to normal. [ABSTRACT FROM AUTHOR]
– Name: AbstractSuppliedCopyright
  Label:
  Group: Ab
  Data: <i>Copyright of Pacing & Clinical Electrophysiology is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.)
PLink https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=17484561
RecordInfo BibRecord:
  BibEntity:
    Identifiers:
      – Type: doi
        Value: 10.1111/j.1540-8159.1988.tb03963.x
    Languages:
      – Code: eng
        Text: English
    PhysicalDescription:
      Pagination:
        PageCount: 11
        StartPage: 1119
    Subjects:
      – SubjectFull: Cardiac pacemakers
        Type: general
      – SubjectFull: Central venous pressure
        Type: general
      – SubjectFull: Blood pressure
        Type: general
      – SubjectFull: Exercise
        Type: general
      – SubjectFull: Fever
        Type: general
      – SubjectFull: Venous pressure
        Type: general
    Titles:
      – TitleFull: A New Rate-Modulated Pacemaker System Optimized by Combination of Two Sensors.
        Type: main
  BibRelationships:
    HasContributorRelationships:
      – PersonEntity:
          Name:
            NameFull: Alt, Eckhard
      – PersonEntity:
          Name:
            NameFull: Theres, Heinz
      – PersonEntity:
          Name:
            NameFull: Heinz, Michael
      – PersonEntity:
          Name:
            NameFull: Matula, Markus
      – PersonEntity:
          Name:
            NameFull: Thilo, Rudolf
      – PersonEntity:
          Name:
            NameFull: Blömer, Hans
    IsPartOfRelationships:
      – BibEntity:
          Dates:
            – D: 01
              M: 08
              Text: Aug1988
              Type: published
              Y: 1988
          Identifiers:
            – Type: issn-print
              Value: 01478389
          Numbering:
            – Type: volume
              Value: 11
            – Type: issue
              Value: 8
          Titles:
            – TitleFull: Pacing & Clinical Electrophysiology
              Type: main