A New Rate-Modulated Pacemaker System Optimized by Combination of Two Sensors.
Saved in:
| Title: | A New Rate-Modulated Pacemaker System Optimized by Combination of Two Sensors. |
|---|---|
| Authors: | Alt, Eckhard1, Theres, Heinz1, Heinz, Michael1, Matula, Markus1, Thilo, Rudolf1, Blömer, Hans1 |
| Source: | Pacing & Clinical Electrophysiology. Aug1988, Vol. 11 Issue 8, p1119-1129. 11p. |
| Subjects: | Cardiac pacemakers, Central venous pressure, Blood pressure, Exercise, Fever, Venous pressure |
| Abstract: | A new rate-modulated pacemaker system optimized by combination of two sensors is described. The parameter body activity and central venous blood temperature control the pacemaker rate. The specific characteristic of each parameter determines its role within the algorithm. While the motion sensor yields a fast reaction following the onset OT a change of stress intensity, central venous blood temperature corresponds better to body metabolism. An indication of increased exercise from the motion sensor results in an accordingly rapid increase in the pacing rate. Unless this increased exercise is confirmed by an increase in central venous blood temperature within 2 or 3 minutes, the new motion level will be assumed to he the new baseline motion value and the pace rate will return to a basic pacing rate. Prolonged inappropriate responses are therefore avoided. Longer lasting exercise, fever and nonphysiological signals are recognized and handled safely. Exercise tests with five volunteers under various conditions showed pacing rate behavior that was close to normal. [ABSTRACT FROM AUTHOR] |
| Copyright of Pacing & Clinical Electrophysiology is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.) | |
| Database: | Engineering Source |
| FullText | Links: – Type: pdflink Text: Availability: 0 |
|---|---|
| Header | DbId: egs DbLabel: Engineering Source An: 17484561 AccessLevel: 6 PubType: Academic Journal PubTypeId: academicJournal PreciseRelevancyScore: 0 |
| IllustrationInfo | |
| Items | – Name: Title Label: Title Group: Ti Data: A New Rate-Modulated Pacemaker System Optimized by Combination of Two Sensors. – Name: Author Label: Authors Group: Au Data: <searchLink fieldCode="AR" term="%22Alt%2C+Eckhard%22">Alt, Eckhard</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Theres%2C+Heinz%22">Theres, Heinz</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Heinz%2C+Michael%22">Heinz, Michael</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Matula%2C+Markus%22">Matula, Markus</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Thilo%2C+Rudolf%22">Thilo, Rudolf</searchLink><relatesTo>1</relatesTo><br /><searchLink fieldCode="AR" term="%22Blömer%2C+Hans%22">Blömer, Hans</searchLink><relatesTo>1</relatesTo> – Name: TitleSource Label: Source Group: Src Data: <searchLink fieldCode="JN" term="%22Pacing+%26+Clinical+Electrophysiology%22">Pacing & Clinical Electrophysiology</searchLink>. Aug1988, Vol. 11 Issue 8, p1119-1129. 11p. – Name: Subject Label: Subjects Group: Su Data: <searchLink fieldCode="DE" term="%22Cardiac+pacemakers%22">Cardiac pacemakers</searchLink><br /><searchLink fieldCode="DE" term="%22Central+venous+pressure%22">Central venous pressure</searchLink><br /><searchLink fieldCode="DE" term="%22Blood+pressure%22">Blood pressure</searchLink><br /><searchLink fieldCode="DE" term="%22Exercise%22">Exercise</searchLink><br /><searchLink fieldCode="DE" term="%22Fever%22">Fever</searchLink><br /><searchLink fieldCode="DE" term="%22Venous+pressure%22">Venous pressure</searchLink> – Name: Abstract Label: Abstract Group: Ab Data: A new rate-modulated pacemaker system optimized by combination of two sensors is described. The parameter body activity and central venous blood temperature control the pacemaker rate. The specific characteristic of each parameter determines its role within the algorithm. While the motion sensor yields a fast reaction following the onset OT a change of stress intensity, central venous blood temperature corresponds better to body metabolism. An indication of increased exercise from the motion sensor results in an accordingly rapid increase in the pacing rate. Unless this increased exercise is confirmed by an increase in central venous blood temperature within 2 or 3 minutes, the new motion level will be assumed to he the new baseline motion value and the pace rate will return to a basic pacing rate. Prolonged inappropriate responses are therefore avoided. Longer lasting exercise, fever and nonphysiological signals are recognized and handled safely. Exercise tests with five volunteers under various conditions showed pacing rate behavior that was close to normal. [ABSTRACT FROM AUTHOR] – Name: AbstractSuppliedCopyright Label: Group: Ab Data: <i>Copyright of Pacing & Clinical Electrophysiology is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract.</i> (Copyright applies to all Abstracts.) |
| PLink | https://search.ebscohost.com/login.aspx?direct=true&site=eds-live&db=egs&AN=17484561 |
| RecordInfo | BibRecord: BibEntity: Identifiers: – Type: doi Value: 10.1111/j.1540-8159.1988.tb03963.x Languages: – Code: eng Text: English PhysicalDescription: Pagination: PageCount: 11 StartPage: 1119 Subjects: – SubjectFull: Cardiac pacemakers Type: general – SubjectFull: Central venous pressure Type: general – SubjectFull: Blood pressure Type: general – SubjectFull: Exercise Type: general – SubjectFull: Fever Type: general – SubjectFull: Venous pressure Type: general Titles: – TitleFull: A New Rate-Modulated Pacemaker System Optimized by Combination of Two Sensors. Type: main BibRelationships: HasContributorRelationships: – PersonEntity: Name: NameFull: Alt, Eckhard – PersonEntity: Name: NameFull: Theres, Heinz – PersonEntity: Name: NameFull: Heinz, Michael – PersonEntity: Name: NameFull: Matula, Markus – PersonEntity: Name: NameFull: Thilo, Rudolf – PersonEntity: Name: NameFull: Blömer, Hans IsPartOfRelationships: – BibEntity: Dates: – D: 01 M: 08 Text: Aug1988 Type: published Y: 1988 Identifiers: – Type: issn-print Value: 01478389 Numbering: – Type: volume Value: 11 – Type: issue Value: 8 Titles: – TitleFull: Pacing & Clinical Electrophysiology Type: main |
| ResultId | 1 |